                     Product Details: 
						                        
 Payment & Shipping Terms: 
							
  | 
                            
| Name: | ECG Similator | Application: | Medical Signal Simulate | 
|---|---|---|---|
| Material: | ABS | Color: | Yellow | 
| Battery Type: | Buil-in Lithium | ||
| Highlight: | Medicine Simulator,Medical Simulation Equipment | 
			    	||
ECG simulator 10 leads ECG machine simulator with 2 type waveform patient monitor simulator build-in battery
Feature of medical simulator:
| 
			 
 1.Rechargeable Lithium battery 
  | 
		
| 
			 
 2.10 pcs multi-functional ECG connector, can be connect with different type ECG machine cable or patient monitor cable 
  | 
		
| 
			 
 3.Battery power indicate light 
  | 
		
| 
			 
 4.Low battery power automatic turn off function, protect the Lithium battery 
  | 
		
 
Specification of simulator:
The simulator can generate 2 types waveform
1.Normal ECG waveform
2.Plus-minus triangular waveform
Dimension: 15x10x3cm
N.W.: 0.35kg
Note:
1.The simulator will automatically generate Normal ECG waveform when connect to power
2.The LED display will be off automatically if without operation after 5 seconds, press any key will be returned to display
Key Description:
There are 4 keys, followed by selection keys, increase key, reducing key, the Enter key, in addition to a set of key combinations
Select Key:
This key is used to select the parameter you want to change; there are 4 LED tubes to display the four codes represent what is displayed,
The 1st LED representative waveform code, LED 2-4 representatives the parameter (2 is the value of one hundred, 3 representatives of ten, 4 representatives of bits)
Operating method described separately for each waveform:
1. Normal ECG waveform:
Heart rate setting range: 10-200bpm; (initial value: 60 bpm)
Signal amplitude fixed.
2. Plus-minus triangular waveform:
Frequency range: 10-300bpm; (initial value: 75 bpm)
Amplitude range: 0.1-4Mv (10: 0.1mV, 400: 4mV);
2 modes, Mode 1: Positive waveform; mode II: negative waveform; Use combinations key to choose.
Note: The signal amplitude settings:
amplitude range: 0.1mV-4mV (10: 0.1mV, 400: 4mV); (initial value: 100)
Please note that this amplitude setting will affect all waveforms.
Please note that the signal amplitude settings do not affect the amplitude of the ECG waveform.
![]()
![]()
Contact Person: Mr. Michael
Tel: 86-15818525640