Ultrasound transducers are the heart of any diagnostic ultrasound system. When image quality degrades, the entire examination suffers — and for hospitals, clinics and imaging centers, a faulty probe means delayed diagnoses and lost scanning time. In this repair case study, we walk through how a Hitachi C251 convex ultrasound transducer that was producing missing and incomplete images was fully restored after a crystal (element) replacement.
The customer, an imaging service provider operating a HITACHI ARIETTA 65 system, reported that the C251 convex probe was displaying incomplete images with missing areas. Instead of a full, uniform field of view, the scan showed blank and dark drop-out zones, which made abdominal and OB/Gyn examinations unreliable and clinically unusable.
Upon inspection and bench testing, our engineers confirmed that the root cause was a failure within the piezoelectric crystal array. When one or more crystal elements stop transmitting and receiving acoustic signals correctly, the system cannot reconstruct a complete image — resulting in exactly the partial, incomplete picture the customer described. Because the housing and lens were otherwise in good condition, a targeted repair was the most cost-effective solution compared with purchasing a brand-new probe.
Our technical team performed a full crystal replacement on the C251 convex probe. The procedure included:
After the repair, the same C251 transducer produced a complete, even and artifact-free image with no drop-outs. The customer confirmed the scan quality was excellent, effectively restoring the probe to full clinical performance. The images below were captured after the crystal replacement.
This repair was carried out on a Hitachi C251 convex ultrasound transducer (Model C251, 1.0-5.0 MHz broadband, convex array, ARIETTA 65 compatible).
If you are dealing with a transducer showing missing, incomplete or degraded images — or any probe fault — send us the model and symptoms and we will provide a fast repair assessment. Contact us today for pricing, spare parts and service partnerships.